Ann Thorac Surg 1992;54:528-532
© 1992 The Society of Thoracic Surgeons
St. Luke's Medical Center, Milwaukee, Wisconsin USA
Accepted for publication February 13, 1992.
* Address reprint requests to Dr Hazelrigg, Cardiac Surgery Associates, Limited, 2901 W Kinnickinnic Pkwy, Suite 311, Milwaukee, WI 53215.
We retrospectively reviewed our last 100 consecutive patients who had an intraaortic balloon pump (IABP) placed through the ascending aorta for postoperative cardiogenic shock. Eighty-one patients survived to have their IABP removed and were evaluated for complications. Complications that may have been related to the transthoracic route of IABP introduction included balloon rupture in 6.2% ([equation]), cerebral vascular accident in 2.5% ([equation]), transient ischemic attack in 1.2% ([equation]), bleeding at the IABP arteriotomy site in 3.7% ([equation]), and mediastinitis in 3.7% ([equation]). Compared with expected rates of development of complications in this high-risk group of patients, it appeared that balloon rupture and mediastinal bleeding were increased because of the transthoracic placement of the IABP. The rates of neurologic events and mediastinal infection do not appear to be increased. Transthoracic IABP placement avoids ischemic problems in the lower extremities and has proved a useful route for IABP introduction.
This article has been cited by other articles:
![]() |
E. C. McGee Jr., P. M. McCarthy, and N. Moazami Temporary Mechanical Circulatory Support Card. Surg. Adult, January 1, 2008; 3(2008): 507 - 534. [Full Text] |
||||
![]() |
N. Moazami and P. M. McCarthy Temporary Circulatory Support Card. Surg. Adult, January 1, 2003; 2(2003): 495 - 520. [Full Text] |
||||
![]() |
R. J.F. Baskett, W. A. Ghali, A. Maitland, and G. M. Hirsch The intraaortic balloon pump in cardiac surgery Ann. Thorac. Surg., October 1, 2002; 74(4): 1276 - 1287. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Santini and A. Mazzucco Transthoracic Intraaortic Counterpulsation: A Simple Method for Balloon Catheter Positioning Ann. Thorac. Surg., September 1, 1997; 64(3): 859 - 860. [Abstract] [Full Text] |
||||
![]() |
J. H. Burack, P. Uceda, and J. N. Cunningham Jr Transthoracic Intraaortic Balloon Pump: A Simplified Technique Ann. Thorac. Surg., July 1, 1996; 62(1): 299 - 301. [Abstract] [Full Text] |
||||
![]() |
F. Beyersdorf, Z. Mitrev, K. Ihnken, W. Schmiedt, K. Sarai, L. Eckel, O. Friesewinkel, G. Matheis, and G. D. Buckberg CONTROLLED LIMB REPERFUSION IN PATIENTS HAVING CARDIAC OPERATIONS J. Thorac. Cardiovasc. Surg., April 1, 1996; 111(4): 873 - 881. [Abstract] [Full Text] |
||||
![]() |
A. L. Moulton Transthoracic balloon pumps To the editor Ann. Thorac. Surg., February 1, 1993; 55(2): 567 - 567. [PDF] |
||||