The Annals of Thoracic Surgery, Vol 53, 706-708, Copyright © 1992 by The Society of Thoracic Surgeons
PM Heerdt, CG Pond, GA Blessios and M Rosenbloom
We have been comparing cardiac output measured with a novel Doppler
pulmonary artery catheter to that measured by thermodilution and aortic
electromagnetometry in cardiac surgical patients. We report here our
observation of a nearly twofold increase in thermodilution cardiac output
after the acute intraoperative onset of tricuspid regurgitation that was
not confirmed by the novel catheter or direct measurement of aortic blood
flow. We conclude that in some patients, acute tricuspid regurgitation may
lessen the reliability of thermodilution cardiac output.
ARTICLES
Inaccuracy of cardiac output by thermodilution during acute tricuspid regurgitation
Division of Cardiothoracic Anesthesiology, Washington University School of Medicine, St. Louis, MO 63110.
This article has been cited by other articles:
![]() |
C. W. Buffington and E. U. M. Nystrom Neither the Accuracy nor the Precision of Thermal Dilution Cardiac Output Measurements Is Altered by Acute Tricuspid Regurgitation in Pigs Anesth. Analg., April 1, 2004; 98(4): 884 - 890. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. M. HOEPER, R. MAIER, J. TONGERS, J. NIEDERMEYER, J. M. HOHLFELD, M. HAMM, and H. FABEL Determination of Cardiac Output by the Fick Method, Thermodilution, and Acetylene Rebreathing in Pulmonary Hypertension Am. J. Respir. Crit. Care Med., August 1, 1999; 160(2): 535 - 541. [Abstract] [Full Text] |
||||
![]() |
E. A Hessel II Review article : What's new in monitoring the coronary surgery patient? Perfusion, July 1, 1992; 7(3): 161 - 194. [Abstract] [PDF] |
||||