ATS
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Akiyama, K.
Right arrow Articles by Koyanagi, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Akiyama, K.
Right arrow Articles by Koyanagi, H.

The Annals of Thoracic Surgery, Vol 46, 73-78, Copyright © 1988 by The Society of Thoracic Surgeons


ARTICLES

Stent creep of porcine bioprosthesis in the mitral position

K Akiyama, O Sawatani, E Imamura, M Endo, A Hashimoto and H Koyanagi
Department of Cardiovascular Surgery, Tokyo Women's Medical College, Japan.

Stent creep, often associated with valve malfunction, is said to play an important role in the long-term performance of a porcine bioprosthesis. We have measured the angle of the stent post showing maximal inward bending (IBA) on 44 mitral porcine bioprostheses. All of them were explanted 1 to 12 years (mean explant time, 7.8 years) postoperatively at reoperation. Patients included 19 men and 25 women, ranging in age from 24 to 66 years (mean age, 47.2 years). Mean IBA was 12.7 +/- 4.2 [SD] degrees in 14 valves implanted for 7 years or less and 16.4 +/- 5.8 degrees in 30 valves implanted 8 years or longer (p less than 0.05). There was no significant difference in IBA among valves based on type (25 Hancock valves, 15.0 +/- 4.7 degrees; 10 Angell-Shiley valves, 16.2 +/- 8.6 degrees; and 9 Carpentier-Edwards valves, 14.4 +/- 3.7 degrees). There was a significant difference in IBA based on valve size (37 valves measuring 25-29 mm in diameter, 16.3 +/- 5.1 degrees; and 7 valves measuring 31 mm in diameter, 9.9 +/- 5.1 degrees; p less than 0.05). IBA showed a tendency to have a large value in a heart with a small left ventricular end-systolic volume. We conclude that (1) stent creep is not related to the materials or designs of the stent post, but tends to increase with passage of time in place; and (2) compression of the left ventricular wall is one of the main causes of stent-post bending.


This article has been cited by other articles:


Home page
J. Thorac. Cardiovasc. Surg.Home page
D. D. Glower, W. D. White, A. C. Hatton, L. R. Smith, W. G. Young, W. G. Wolfe, and J. E. Lowe
Determinants of reoperation after 960 valve replacements with Carpentier-Edwards prostheses