W Klepetko, A Moritz, G Khunl-Brady, W Schreiner, W Schlick, J Mlczoch, G Kronik and E Wolner
From September, 1983, to April, 1986, 451 Duromedics bileaflet cardiac
valve prostheses were implanted in 400 patients at our institution in
Vienna. Aortic valve replacement was done in 190 patients, 157 underwent
mitral valve replacement (1 patient also underwent tricuspid valve
replacement), 52 underwent double valve replacement, and 1 patient
underwent isolated reoperation for tricuspid valve replacement. Concomitant
procedures were performed in 86 patients (21.5%). Sixty-one patients
(15.2%) had undergone previous cardiac surgery; 32 (8%) had undergone
earlier valve replacement. The early mortality rate (within 30 days) was
6.25% (25 patients). Follow-up was done on 337 surviving Austrian citizens;
this represents 429 patient-years. The late mortality rate was 2.1% per
patient-year (9 patients). We observed paravalvular leak in 3 patients
(0.7% per patient-year), thromboembolism in 4 (0.9%), prosthetic valve
endocarditis in 5 (1.2%), and anticoagulant-related hemorrhage in 10
(2.3%). Valve failure occurred in 8 patients (1.8%). We conclude,
therefore, that good clinical results and a low complication rate can be
achieved with this new type of valve.
ARTICLES
Implantation of the Duromedics bileaflet cardiac valve prosthesis in 400 patients
This article has been cited by other articles:
![]() |
N. D. Desai and G. T. Christakis Bioprosthetic Aortic Valve Replacement: Stented Pericardial and Porcine Valves Card. Surg. Adult, January 1, 2008; 3(2008): 857 - 894. [Full Text] |
||||
![]() |
V. A. Ferraris and S. P. Ferraris Assessing the medical literature: let the buyer beware Ann. Thorac. Surg., July 1, 2003; 76(1): 4 - 11. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. D. Desai and G. T. Christakis Stented Mechanical/Bioprosthetic Aortic Valve Replacement Card. Surg. Adult, January 1, 2003; 2(2003): 825 - 856. [Full Text] |
||||
![]() |
C. E.E. Kuntze, E. H. Blackstone, and T. Ebels Thromboembolism and mechanical heart valves: a randomized study revisited Ann. Thorac. Surg., July 1, 1998; 66(1): 101 - 107. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. A. Blome-Eberwein, D. Mrowinski, J. Hofmeister, and R. Hetzer Impact of Mechanical Heart Valve Prosthesis Sound on Patients' Quality of Life Ann. Thorac. Surg., February 1, 1996; 61(2): 594 - 602. [Abstract] [Full Text] |
||||
![]() |
H. E. Scully and C. S. Armstrong Tricuspid valve replacement: Fifteen years of experience with mechanical prostheses and bioprostheses J. Thorac. Cardiovasc. Surg., June 1, 1995; 109(6): 1035 - 1041. [Abstract] [Full Text] |
||||