The Annals of Thoracic Surgery, Vol 42, 697-698, Copyright © 1986 by The Society of Thoracic Surgeons
E Milgalter, M Mosseri, G Uretzky and H Romanoff
A patient in whom perforation of a percutaneously inserted intraaortic
balloon resulted in a clot formation inside the balloon is described. The
balloon could not be withdrawn percutaneously and was lodged in the femoral
artery. It was removed surgically, and the artery was repaired.
ARTICLES
Intraaortic balloon entrapment: a complication of balloon perforation
This article has been cited by other articles:
![]() |
C. L. Chiu, M. Mansor, A. Majid, C. Marcucci, P. G. Chassot, and D. R. Spahn Anaesthetic management of high-risk cardiac patients undergoing thoracic surgery with the support of intra-aortic balloon pump Br. J. Anaesth., May 1, 2005; 94(5): 688 - 689. [Full Text] [PDF] |
||||
![]() |
L. M. Harris, R. L. Guarino, L. Jennings, and J. J. Ricotta Intraaortic Balloon Pump Entrapment: An Alternative Method of Management: A Case Report Vascular and Endovascular Surgery, May 1, 1996; 30(3): 239 - 244. [Abstract] [PDF] |
||||