The Annals of Thoracic Surgery, Vol 42, 74-76, Copyright © 1986 by The Society of Thoracic Surgeons
M Galinanes, J Duarte, DF de Caleya, D Garcia-Dorado, F Fdez-Aviles and LM Elbal
Three cases of fracture of a Carpentier-Edwards ring in tricuspid position
are presented. In all three cases the shape of the ring was shown to be
abnormal by roentgenogram examination; there was an abrupt bend in the
middle portion of each ring. At the time of operation, fracture of the ring
skeleton at this midpoint was found in all three instances. Roentgenogram
examination seems to be a reliable method of diagnosing this uncommon
entity, and tricuspid valve replacement is required for its correction.
ARTICLES
Fracture of the Carpentier-Edwards ring in tricuspid position: a report of three cases
This article has been cited by other articles:
![]() |
M. Hartrumpf, U. A. Stock, R.-U. Kuehnel, and J. M. Albes Fracture of colvin-galloway future band causing a tear in the anterior mitral leaflet. Ann. Thorac. Surg., May 1, 2006; 81(5): 1879 - 1880. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |