The Annals of Thoracic Surgery, Vol 32, 50-57, Copyright © 1981 by The Society of Thoracic Surgeons
HV Schaff, AM Borkon, C Hughes, S Achuff, JS Donahoo, TJ Gardner, L Watkins Jr, VL Gott, AG Morrow and RK Brawley
Between November, 1973, and March, 1980, 43 patients underwent isolated
aortic valve replacement with 19 mm Bjork-Shiley prostheses at the Johns
Hopkins Hospital. There were 4 male and 39 female patients ranging from 12
to 75 years old (mean, 54.5 years). Average weight was 62 +/- 2 kg and
average body surface area, 1.64 +/- 0.3 m2. Five patients died within
thirty days of operation; however, since 1975, hospital mortality has been
5.9%. The 38 survivors have been followed up for as long as 85 months
(mean, 40 months). There were 4 late deaths, and actuarial survival in
patients discharged from the hospital was 81% at five years. All long-term
survivors were in New York Heart Association Functional Class I (29
patients) or Class II (5 patients). Preoperative and postoperative
echocardiograms in 17 patients demonstrated significant decreases in mean
left ventricular wall thickness (12.9 +/- 1.8 mm vs 10.3 +/- 1.4 mm; p less
than 0.001) and in left ventricular mass (262 +/- 95 gm vs 188 +/- 50 gm; p
less than 0.02). Postoperative cardiac catheterization data were obtained
from an additional 24 patients undergoing aortic valve replacement with the
19 mm Bjork-Shiley prosthesis at the National Heart Institute. Average peak
systolic gradient at rest was 16 mm Hg (range, 0 to 45 mm Hg) and was found
to be directly related to body surface area (r = 0.60, p less than 0.002).
Average effective valve orifice area was 1.06 cm2 (range, 0.63 to 2.02
cm2). For patients with small aortic roots, aortic valve replacement with
the 19 mm Bjork-Shiley valve is a satisfactory and, perhaps, preferable
alternative to aortic annuloplasty to accommodate larger sized prostheses.
ARTICLES
Clinical and hemodynamic evaluation of the 19 mm Bjork-Shiley aortic valve prosthesis
This article has been cited by other articles:
![]() |
H. Takakura, T. Sasaki, K. Hashimoto, T. Hachiya, K. Onoguchi, M. Oshiumi, and S. Takeuchi Hemodynamic evaluation of 19-mm Carpentier-Edwards pericardial bioprosthesis in aortic position Ann. Thorac. Surg., February 1, 2001; 71(2): 609 - 613. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. B. Izzat, I. Kadir, B. Reeves, P. Wilde, A. J. Bryan, and G. D. Angelini Patient-prosthesis mismatch is negligible with modern small-size aortic valve prostheses Ann. Thorac. Surg., November 1, 1999; 68(5): 1657 - 1660. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. De Paulis, L. Sommariva, G. M. De Matteis, E. Caprara, F. Tomai, A. P. de Peppo, P. Polisca, C. Bassano, and L. Chiariello EXTENT AND PATTERN OF REGRESSION OF LEFT VENTRICULAR HYPERTROPHY IN PATIENTS WITH SMALL SIZE CARBOMEDICS AORTIC VALVES J. Thorac. Cardiovasc. Surg., May 1, 1997; 113(5): 901 - 909. [Abstract] [Full Text] |
||||
![]() |
D. Sawant, A. K. Singh, W. C. Feng, A. A. Bert, and F. Rotenberg Nineteen-Millimeter Aortic St. Jude Medical Heart Valve Prosthesis: Up to Sixteen Years' Follow-up Ann. Thorac. Surg., April 1, 1997; 63(4): 964 - 970. [Abstract] [Full Text] |
||||
![]() |
D. Sawant, A. K. Singh, W. C. Feng, A. A. Bert, and F. Rotenberg ST. JUDE MEDICAL CARDIAC VALVES IN SMALL AORTIC ROOTS: FOLLOW-UP TO SIXTEEN YEARS J. Thorac. Cardiovasc. Surg., March 1, 1997; 113(3): 499 - 509. [Abstract] [Full Text] |
||||
![]() |
M. B. Izzat, I. Birdi, P. Wilde, A. J. Bryan, and G. D. Angelini COMPARISON OF HEMODYNAMIC PERFORMANCES OF ST. JUDE MEDICAL AND CARBOMEDICS 21 MM AORTIC PROSTHESES BY MEANS OF DOBUTAMINE STRESS ECHOCARDIOGRAPHY J. Thorac. Cardiovasc. Surg., February 1, 1996; 111(2): 408 - 415. [Abstract] [Full Text] |
||||
![]() |
T. Ota, K. Iwahashi, H. Matsuda, T. Tsukube, K. Ataka, and M. Okada Reduction of Left Ventricular Hypertrophy with St. Jude Medical 19 mm Valve Prosthesis Angiology, November 1, 1995; 46(11): 981 - 987. [Abstract] [PDF] |
||||
![]() |
M. B. Izzat, I. Birdi, P. Wilde, A. J. Bryan, and G. D. Angelini Evaluation of the Hemodynamic Performance of Small CarboMedics Aortic Prostheses Using Dobutamine-Stress Doppler Echocardiography Ann. Thorac. Surg., October 1, 1995; 60(4): 1048 - 1052. [Abstract] [Full Text] |
||||
![]() |
G.-W. He, G. L. Grunkemeier, H. L. Gately, A. P. Furnary, and A. Starr Up to Thirty-Year Survival After Aortic Valve Replacement in the Small Aortic Root Ann. Thorac. Surg., May 1, 1995; 59(5): 1056 - 1062. [Abstract] [Full Text] |
||||
![]() |
R. De Paulis, L. Sommariva, F. Russo, F. Tomai, A. Tondo, C. Pagliaricci, C. Bassano, and L. Chiariello Doppler echocardiography evaluation of the CarboMedics valve in patients with small aortic anulus and valve prosthesis-body surface area mismatch J. Thorac. Cardiovasc. Surg., July 1, 1994; 108(1): 57 - 62. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |