Ann Thorac Surg 1980;29:381-384
© 1980 The Society of Thoracic Surgeons
Division of Cardiothoracic Surgery, Department of Surgery, and Nursing Services, University of Washington School of Medicine, Seattle WA
Accepted for publication May 4, 1979.
* Address reprint requests to Dr. Hessel, Division of Cardiothoracic Surgery, Department of Surgery, University of Washington School of Medicine, BB 442, RF-25, Seattle, WA 98195
Intraaortic balloon pumping is widely utilized for mechanical circulatory support of patients with left ventricular power failure. Although intraaortic balloon pump consoles are quite reliable, malfunctions do occur. If a standby console of similar make is not available, a console of a different manufacturer can be utilized to continue counterpulsation. This article provides guidelines for the emergency driving of a Datascope pump console and an AVCO balloon catheter, and an AVCO pump console and a Datascope balloon catheter. The successful use of the Datascope pump and an AVCO balloon in 5 patients is reported.
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |