Ann Thorac Surg 1978;26:427-437
© 1978 The Society of Thoracic Surgeons
Cardiovascular Surgical Research Laboratory, Children's Hospital Medical Center, and the Department of Surgery, Harvard Medical School, Boston, MA.
* Address reprint requests to Dr. Bernhard, Department of Surgery, Children's Hospital Medical Center, 300 Longwood Ave, Boston, MA 02115.
Mechanical circulatory support was accomplished in 20 calves (mean, 140 days) and in 5 patients following operation for acquired heart disease (range, 1 hour to 8 days) employing a pneumatically actuated xenograft-valved assist pump interposed between the left ventricular apex and aorta.
Following pump implantation in calves, hematocrit and platelets decreased transiently and returned to normal within 14 days. Plasma hemoglobin and erythrocyte mechanical fragility values were elevated for 48 hours. Platelet survival was slightly reduced, but erythrocyte survival values were similar to controls. In patients who received assist pumps, plasma hemoglobin and erythrocyte mechanical fragility were transiently elevated, but rapidly decreased to normal. Thrombocytopenia occurred only in the presence of bleeding and renal failure requiring hemodialysis. Pump flow of the left ventricular assist device was maintained above 2.0 L/min/m2 despite serious arrhythmias. Postmortem examination revealed no evidence of thromboemboli in the clinical patients although anticoagulant agents were not administered.
This article has been cited by other articles:
![]() |
R. J. Levy, J. A. Zenker, and W. F. Bernhard Porcine Bioprosthetic Valve Calcification in Bovine Left Ventricle-Aorta Shunts: Studies of the Deposition of Vitamin K-Dependent Proteins Ann. Thorac. Surg., August 1, 1983; 36(2): 187 - 192. [Abstract] [PDF] |
||||
![]() |
J. C. Pollock, M. C. Charlton, W. G. Williams, J. F. Edmonds, and G. A. Trusler Intraaortic Balloon Pumping in Children Ann. Thorac. Surg., June 1, 1980; 29(6): 522 - 528. [Abstract] [PDF] |
||||
![]() |
W. Pierce, J. Donachy, G Rosenberg, and R. Baier Calcification inside artificial hearts: inhibition by warfarin-sodium Science, May 9, 1980; 208(4444): 601 - 603. [Abstract] [PDF] |
||||
![]() |
R. L. Berger, J. R. McCormick, J. D. Stetz, M. D. Klein, T. J. Ryan, J. Carr, S. Sweet, and W. F. Bernhard Successful Use of a Paracorporeal Left Ventricular Assist Device in Man JAMA, January 4, 1980; 243(1): 46 - 49. [Abstract] [PDF] |
||||
![]() |
L. R. Golding, L. K. Groves, M. Peter, G. Jacobs, R. Sukalac, Y. Nose, and F. D. Loop Initial Clinical Experience with a New Temporary Left Ventricular Assist Device Ann. Thorac. Surg., January 1, 1980; 29(1): 66 - 69. [Abstract] [PDF] |
||||