The Annals of Thoracic Surgery, Vol 23, 45-51, Copyright © 1977 by The Society of Thoracic Surgeons
PA Philips and D Bregman
Persistent unrecognized subendocardial ischemia with development of
subendocardial necrosis is a major cause of patient death following
cardiopulmonary bypass. The lesion is caused by a discrepancy between the
oxygen needs of subendocardial muscle and the available blood supply. If
sole reliance is placed upon monitoring conventional vital signs, the more
subtle factors contributing to decreased blood flow may go unrecognized.
Reported studies have confirmed that the adequacy of subendocardial
perfusion can be predicted by calculating the supply/demand ratio, defined
as the ratio of the diastolic pressure- time index (DPTI) divided by the
systolic pressure-time index (TTI). An analog computer was designed and
built that measures the area under the systolic and diastolic component,
calculates the DPTI/TTI ratio, and digitally displays the result as the
endocardial viability ratio (evr). The EVR was used to determine the
adequacy of left ventricular subendocardial blood flow in 64 consecutive
patients undergoing cardiac operations. Unidirectional intraaortic balloon
counterpulsation (IABC) was utilized in 14 patients with 9 long-term
survivors. The difference in mean EVR between survivors and nonsurvivors at
the initiation of balloon support was statistically significant. Early
application of unidirectional IABC when subendocardial ischemia persists
following open cardiac procedures may prevent deterioration to
subendocardial necrosis with subsequent morbidity or mortality.
ARTICLES
Intraoperative application of intraaortic balloon counterpulsation determined by clinical monitoring of the endocardial viability ratio
This article has been cited by other articles:
![]() |
J. N. Nanas, M. Anastasiou-Nana, and S. D. Moulopoulos Experimental and Clinical Evaluation of a Counterpulsation Device J Biomater Appl, April 1, 1990; 4(4): 362 - 373. [Abstract] |
||||
![]() |
J. Chalmers and T. Graham Review article : Mechanical ventricular support in the management of postcardiotomy cardiogenic shock Perfusion, October 1, 1989; 4(4): 243 - 253. [Abstract] [PDF] |
||||
![]() |
J. M. Smith, L. D. Joyce, H. G. Mauer, and R. C. Lillehei Intraaortic Balloon Counterpulsation: A Review Vascular and Endovascular Surgery, November 1, 1978; 12(6): 377 - 384. [Abstract] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |