The Annals of Thoracic Surgery, Vol 19, 301-308, Copyright © 1975 by The Society of Thoracic Surgeons
DA Hughes, SR Igo, BD Daly, JJ Migliore and JC Norman
The effects of an abdominal left ventricular assist device (ALVAD) on
subendocardial oxygen supply/demand ratios as reflected by diastolic
pressure-time index/tension-time index (DPTI/TTI) were investigated in
calves. Control DPTI/TTI ratios were 1.09 plus or minus 0.12 (mean plus or
minus SD). With ALVAD actuation the ratios increased to 5.73 plus or minus
3.2. Following circumflex coronary artery occlusion, control ratios
decreased to 0.04 plus or minus 0.05 in the ischemic area and 0.97 plus or
minus 0.20 in the nonischemic areas. With ALVAD actuation in the presence
of circumflex occlusion, the depressed ratios were increased to 1.37 plus
or minus 0.39 in the ischemic zone and 9.71 plus or minus 2.5 in the
nonoccluded segments. Subendocardial oxygen supply/demand ratios were
significantly improved by ALVAD assistance following induced subendocardial
ischemia.
ARTICLES
Effects of an abdominal left ventricular assist device on myocardial oxygen supply/demand ratios in normally perfused and ischemic bovine myocardium
This article has been cited by other articles:
![]() |
M. Szycher, W. Clay, D. Gernes, and C. Sherman Thermedics' Approach to Ventricular Support Systems J Biomater Appl, January 1, 1986; 1(1): 39 - 105. [Abstract] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |